Influence of mandibular positioning on the accuracy and reliability of linear measurements on cone beam CT using different voxels and two software programs
DOI:
https://doi.org/10.5335/rfo.v25i3.13336Keywords:
dimensional measurement accuracy; cone beam computed tomography; diagnostic imaging; mandible; software.Abstract
Objective: this study evaluated the accuracy and reliability of linear measurements on cone beam CT (CBCT)
scans in two software programs, using different voxels and varying mandible positioning. Material and methods:
CBCT images of 10 human mandibles with 25 markers were obtained using different acquisition protocols
(0.250, 0.300, 0.400-mm voxels) and mandible orientations (centered, rotated 10° laterally to right and
left, tilted 10 up and down); fourteen measurements were carried out on the multiplanar reconstructions in
XoranCat and OsiriX Lite software programs. The findings were compared to physical measurements using a
digital caliper. ANOVA and correlation coefficient tests were used, at α = 0.05. Results: there was no statistically
significant difference when the measurements were compared in acquisitions with different voxel sizes
analysed in both software programs. Mandibular positioning changes did not influence the measurements.
No differences were found when the values were compared between the software programs and the digital
caliper. Conclusion: linear measurements in both programs were reliable and accurate compared with physical
measurements when using all acquisition protocols. The accuracy and reliability of the measurements
were not influenced by variations in the mandible positioning.
Downloads
References
Dental Clinics of North America 2018; 62(3):361-91.
2. De Vos W, Casselman J, Swennen GR. Cone-beam computerized
tomography (CBCT) imaging of the oral and maxillofacial
region: a systematic review of the literature. Int J Oral
Maxillofac Surg 2009; 38:609-25.
3. Schulze RKW, Drage NA. Cone-beam computed tomography
and its applications in dental and maxillofacial radiology.
Clinical Radiology 2020; 75(9):647-57.
4. Ludlow JB, Ivanovic M. Comparative dosimetry of dental
CBCT devices and 64-slice CT for oral and maxillofacial radiology.
Oral Surgery, Oral Medicine, Oral Pathology, Oral
Radiology, and Endodontology 2008; 106(1):106-14.
5. Caetano APF, Sousa TO, Oliveira MR, Evanglista K, Bueno
JM, Silva MA. Accuracy of three cone-beam CT devices and
two software systems in the detection of vertical root fractures.
Dentomaxillofac Radiol 2021; 50(3):20200334.
6. Vaz SLA, Vasconcelos TV, Neves FS, Freitas DQ, Haiter-Neto
F. Influence of cone beam computed tomography enhancement
filters on diagnosis of simulated external root resorption.
J Endod 2012; 38:305-8.
7. Lascala C, Panelaa J, Marques M. Analysis of the accuracy
of linear measurements obtained by cone beam computed
tomography (CBCT-NewTom). Dentomaxillofac Radiol 2004;
33:291-4.
8. Baumgaertel S, Palomo JM, Palomo L, Hans MG. Reliability
and accuracy of cone-beam computed tomography dental
measurements. Am J Orthod Dentofacial Orthop 2009;
136:19-25.
9. Tolentino ES, Yamashita FC, De Albuquerque S, Walewski
LA, Iwaki LC, Takeshita WM, et al. Reliability and accuracy
of linear measurements in cone-beam computed tomography
using different software programs and voxel sizes. J Conserv
Dent 2018; 21:607-12.
10. Melo SLS, Haiter-Neto F, Correa LR, Scarfe WC, Farman
AG. Comparative diagnostic yield of cone beam CT reconstruction
using various software programs on the detection
of vertical root fractures. Dentomaxillofac Radiol 2013;
42:20120459.
11. Costa FF, Gaiz BF, Umersubo OS, Pinheiro LR, Tortamo IP,
Cavalcanti MGP. Use of large-volume cone-beam computed
tomography in identification and localization of horizontal
root fracture in the presence and absence of intracanal metallic
post. J Endod 2012; 38:856-9.
12. Hassan B, Setelt P, Sanderink G. Accuracy of three-dimensional
measurements obtained from cone beam computed
tomography surface-rendered images for cephalometric
analysis: influence of patient scanning position. Eur J Orthod
2008; 23:1-6.
13. El-Beialy AR, Fayed MS, El-Bialy AM, Mostafad YA. Accuracy
and reliability of cone-beam computed tomography
measurements: Influence of head orientation. Am J Orthod
Dentofacial Orthop 2011; 140:140-57.
14. Librizzi ZT, Tadinada AS, Valiyaparambil JV, Lurie AG,
Mallya SM. Cone-beam computed tomography to detect erosions
of the temporomandibular joint: effect of field of view
and voxel size on diagnostic efficacy and effective dose. Am J
Orthod Dentofacial Orthop 2011;140:e25-30.
15. Poleti ML, Fernandes TM, Pagin O, Moretti MR, Rubira-
-Bullen IR. Analysis of linear measurements on 3D surface
models using CBCT data segmentation obtained by automatic
standard pre-set thresholds in two segmentation software
programs: an in vitro study. Clin Oral Investig 2016; 20:179-
85.
16. Grauer D, Cevidanes LSH, Proffit WR. Working with DICOM
craniofacial images. Am J Orthod Dentofacial Orthop
2009; 136:460-70.
17. Periago DR, Scarfe WC, Moshiri M, Scheetz JP, Silveira AM,
Farman AG. Linear accuracy and reliability of cone beam CT
derived 3-dimensional images constructed using an orthodontic
volumetric rendering program. Angle Orthod 2008;
78:387-95.
18. Ballrick JW, Palomo JM, Ruch E, Amberman BD, Hans MG.
Image distortion and spatial resolution of a commercially
available cone-beam computed tomography machine. Am J
Orthod Dentofacial Orthop 2008; 134:573-82.
19. Timock AM, Cook V, McDonald T, Leo MC, Crowe J, Benninger
BL, et al. Accuracy and reliability of buccal bone height
and thickness measurements from cone-beam computed tomography
imaging. Am J Orthod Dentofacial Orthop 2011;
140:734-44.
20. El H, Palomo JM. Measuring the airway in 3 dimensions:
a reliability and accuracy study. Am J Orthod Dentofacial
Orthop 2010; 137: S50.e1–S50.e9.
21. Gungor E, Dogan MS. Reliability and accuracy of cone-beam
computed tomography voxel density and linear distance
measurement at different voxel sizes: A study on sheep head
cadaver. J Dent Sci 2017; 12(2):145-50.
22. Liedke GS, Silveira HED, Silveira HLD, Dutra V, Figueiredo
JAP. Influence of voxel size in the diagnostic ability of
cone beam tomography to evaluate simulated external root
resorption. J Endod 2009; 35:233-5.
23. Berco M, Rigali PHJR, Miner MR, Deluca S, Anderson NK,
Will LA. Accuracy and reliability of linear cephalometric
measurements from cone-beam computed tomography scans
of a dry human skull. Am J Orthod Dentofacial Orthop 2009;
136:17.e1-9.
Downloads
Published
Issue
Section
License

This work is licensed under aCreative Commons Atribuição-NãoComercial-SemDerivações 4.0 Internacional.